RAZOR™ Nano Chamber, with the smallest volume in the market and high performance physics characteristics, is a unique ionization chamber solution for small field dosimetry.
Compatible with: Small Field Dosimetry, C-arm linacs
Used for: Commissioning and Annual QA RT, Detectors RT
Features
- Cavity volume of 0.003 ccm
- Central electrode material: graphite
- For photon and electron beams in RT
Technical specifications
- PFD³G Photon — Effective measurement point: < 0.9 mm; Chip size (side/thickness): 2.5 / 0.5 mm; Geometric form of active area: circled; Diameter of active area: 2 mm; Thickness of active volume: 0.06 mm
- EFD³G Electron — Effective measurement point: < 0.9 mm; Chip size (side/thickness): 2.5 / 0.5 mm; Geometric form of active area: circled; Diameter of active area: 2 mm; Thickness of active volume: 0.06 mm
- RFD³G Reference — Effective measurement point: n. a.; Chip size (side/thickness): 2.5 / 0.5 mm; Geometric form of active area: circled; Diameter of active area: 2 mm; Thickness of active volume: 0.06 mm
- RAZOR DIODE — Effective measurement point: 0.8 ± 0.2 mm; Chip size (side/thickness): 0.95 / 0.4 mm; Geometric form of active area: circled; Diameter of active area: 0.6 mm; Thickness of active volume: 0.02 mm