PPC40 is a parallel plate ionization chamber for absolute and relative dosimetry of electron (≥2 MeV) and proton beams, as well as photon depth-dose measurements. Made entirely of PMMA with graphitized electrodes, PPC40 features a circular planar-sensitive volume with a rigid 1 mm entrance window.

PPC40 Plane Parallel Chamber

PPC40 is a parallel plate ionization chamber for absolute and relative dosimetry of electron (≥2 MeV) and proton beams, as well as photon depth-dose measurements. Made entirely of PMMA with graphitized electrodes, PPC40 features a circular planar-sensitive volume with a rigid 1 mm entrance window. Waterproof with a ventilated silicone-covered cable, it is ideal for precise absolute dose determinations in reference dosimetry and equipment setup.

Used for: Commissioning PT, Detectors RT, Monthly QA RT, Monthly QA PT, Annual QA PT, Commissioning and Annual QA RT

Features

Outer dimensions

  • Chamber outer diameter: 44.0 mm
  • Chamber body height: 10.0 mm
  • Stem diameter: 10.0 mm
  • Stem length: 50.0 mm

Inner dimensions

  • Sensitive volume (nominal): 0.40 cm³
  • Cylinder height: 2.0 mm
  • Entrance window (polarizing electrode) diameter: 20.0 mm
  • Entrance window thickness: 1.0 mm
  • Diameter of collecting electrode: 16.0 mm
  • Guard ring diameter: 17.8 mm
  • Guard ring width: 3.4 mm

Operational characteristics

  • Polarizing voltage: ±300 V (max. ±500 V)
  • Typical leakage current: 5 fA
  • Recommended pre-irradiation: 5 Gy
  • Typical sensitivity: 11 nC/Gy
  • Guard potential: ±300 V (max. ±500 V)
  • Temperature range: 15 to 35 °C
  • Relative humidity range: 20 % to 80 %

Technical specifications

  • PPC05 — Materials: Window and body C552; graphitized (PEEK) electrode; Window thickness [mg/cm²]/[mm]: 176 / 1; Active volume [cm³]: 0.05; Electrode spacing [mm]: 0.5; Collecting electrode diameter [mm]: 10; Guard ring width [mm]: 3.5; Waterproof: Y
  • PPC40 — Materials: PMMA; Window thickness [mg/cm²]/[mm]: 118 / 1; Active volume [cm³]: 0.4; Electrode spacing [mm]: 2; Collecting electrode diameter [mm]: 16; Guard ring width [mm]: 4; Waterproof: Y
  • NACP — Materials: Mylar foil and graphite window; body PMMA; electrode graphitized; Window thickness [mg/cm²]/[mm]: 104 / 0.6; Active volume [cm³]: 0.16; Electrode spacing [mm]: 2; Collecting electrode diameter [mm]: 10; Guard ring width [mm]: 3; Waterproof: Y

Technical data is subject to change without prior notice.

In this division

DOSE-X

Next-Generation Reference Class Electrometer DOSE-X is a portable reference class electrometer exceeding IEC 60731 standards. Discover your perfect solution for conducting reference measurements according to DIN 6800, AAPM TG51, and IAEA TRS398. Compatible with Machine/Modality/Application: Carbon Compatible , C-arm linacs , Robotic linac , Cyclotrons , Bore-type linacs , IMRT / VMAT , Proton Therapy , Helical linac , Small Field Dosimetry , Synchro-Cyclotron , Ultra-high Dose Rate (UHDR) , TG106, TG142, TG148, TG224, TG51 Used for Workflow Commissioning PT, Commissioning and Annual QA RT, Annual QA PT, Monthly QA RT, Monthly QA PT