Skip to content

Quality Assurance & Dosimetry

Lynx PT

High accuracy and speed for daily machine parameter verification Daily machine parameter verification used from commissioning to machine QA, without compromising between high accuracy and speed.

Lynx PT

High accuracy and speed for daily machine parameter verification Daily machine parameter verification used from commissioning to machine QA, without compromising between high accuracy and speed.

Compatible with: Carbon Compatible, Cyclotrons, Synchro-Cyclotron, Proton Therapy, TG224

Used for: Commissioning PT, Annual QA PT, Monthly QA PT

What you need to know about Lynx PT

Technology

The Lynx detector consists of a box composed by a scintillating screen of gadolinium-based plastic material which converts the energy lost by ionizing radiation into photons. This light is reflected to photodiodes of a CCD camera outside the irradiation field. To avoid the saturation of the camera, iris aperture can be changed, as well as the gain. The detector signal from the CCD camera is digitalized at a 10-bit depth.

The Lynx detector is easy to setup. It can be operated on the treatment couch. Data is transferred to a PC via Ethernet (previous version used Firewire).

Lynx is supported by the following software solutions:

  • Lynx2D
  • myQA Lynx plug-in

Technical specifications

  • Scintillation screen thickness: 0.4 mm
  • Energy range: 60Co at 25 MV in photons /4 to 25 MeV in electrons/ 230 MeV Protons
  • Active surface area: 300 x 300 mm² (600 x 600 pixels)
  • Pixel size: 4.65 x 4.65 um²
  • Effective spacial resolution: 0.5 mm
  • Geometric distortion: ≤ 0.3 mm for the central zone of ≤ 280 x 280 mm²; ≤ 0.8 mm elsewhere
  • Image acquisition: CCD camera
  • Video camera: 1024 x 1024, 12-bit resolution
  • Sampling time: 7 images / secondes
  • Dose linearity within +/- 1.5%
  • Spot sigma: Variation < 100 um
  • Dimensions: 360 x 370 x 600 mm

Accessories

  • Calibration plate for Lynx PDR1602

Sphinx with Lynx Publication

Placidi et.al. “ Range Resolution and Reproducibility of a Dedicated Phantom for Proton PBS Daily Quality Assurance .” Zeitschrift Für Medizinische Physik 28, no. 4 (December 1, 2018): 310–17.

Conclusion The proposed combination of the Lynx ® + Sphinx ® systems appear to be suitable as a device for daily QA of proton range for PBS proton therapy, even under extreme, and very unlikely, set-up and beam delivery uncertainties.

In this division

DOSE-X

Next-Generation Reference Class Electrometer DOSE-X is a portable reference class electrometer exceeding IEC 60731 standards. Discover your perfect solution for conducting reference measurements according to DIN 6800, AAPM TG51, and IAEA TRS398. Compatible with Machine/Modality/Application: Carbon Compatible , C-arm linacs , Robotic linac , Cyclotrons , Bore-type linacs , IMRT / VMAT , Proton Therapy , Helical linac , Small Field Dosimetry , Synchro-Cyclotron , Ultra-high Dose Rate (UHDR) , TG106, TG142, TG148, TG224, TG51 Used for Workflow Commissioning PT, Commissioning and Annual QA RT, Annual QA PT, Monthly QA RT, Monthly QA PT

Talk to our team

Questions about a system, a quotation or after-sales support? Call or email us and the right person will get back to you.