Skip to content

Quality Assurance & Dosimetry

Giraffe

Giraffe is a high resolution multi-layer ionization chamber, which consists of 180 large 12 cm diameter electrodes.

Giraffe

Giraffe is a high resolution multi-layer ionization chamber, which consists of 180 large 12 cm diameter electrodes.

Compatible with: Synchro-Cyclotron, Cyclotrons, Proton Therapy, TG224

Used for: Annual QA PT, Monthly QA PT

Rapid Particle Therapy Commissioning

  • Instant absolute range verification for weekly & monthly QA
  • 3x faster commissioning of your particle therapy rooms
  • Reconstruct the Bragg peak for a single spot within seconds
  • Verify the clinical range of up to 230 MeV

Instant PBS verification within one second

  • Check your complete Bragg peak depth dose curve for single spot beam delivery

Minimize loss of data collection

  • Save time with reduced effort to correct data at TPS level. Verify the TPS library periodically without needing to set up the water tank.

Technical Specifications

  • Application: measurement of depth dose distribution in particle therapy
  • Beam type: single spot pristine and Bragg peak
  • Chamber type: vented ionization chambers
  • Chamber size: 12 cm diameter
  • Range accuracy: ± 0.5 mm WET
  • Energy range: 33 cm WET
  • Spatial resolution: 2 mm detector spacing (native resolution)
  • Dose rate range: 0.5 Gy/min to 15 Gy/min
  • Signal to noise ratio: max. 0.2 % with 1 cGy integrated dose
  • Dose linearity: max. 0.5 % from 10 cGy up to 5 Gy integral dose
  • max. 1 % from 0.5 Gy/min up to 15 Gy/min dose rate
  • Typical sensitivity: 14.76 nC/Gy
  • Electrometer: 4 TERA ASICs (each contains 64 independent electrometers)
  • Channels: 180
  • Charge resolution: 0.1 pC/count
  • Sampling time: min. 10 ms
  • Readout: parallel and synchronous with no dead time
  • Weight: approx. 10 kg
  • Dimensions: 43.9 cm (L) x 19.5 cm (H) x 17.5 cm (W)
  • Power supply: 100-240 V, 50/60 Hz, power cord with US or German power plug included
  • Network: point to point or Ethernet connection

Technical data is subject to change without prior notice.

In this division

DOSE-X

Next-Generation Reference Class Electrometer DOSE-X is a portable reference class electrometer exceeding IEC 60731 standards. Discover your perfect solution for conducting reference measurements according to DIN 6800, AAPM TG51, and IAEA TRS398. Compatible with Machine/Modality/Application: Carbon Compatible , C-arm linacs , Robotic linac , Cyclotrons , Bore-type linacs , IMRT / VMAT , Proton Therapy , Helical linac , Small Field Dosimetry , Synchro-Cyclotron , Ultra-high Dose Rate (UHDR) , TG106, TG142, TG148, TG224, TG51 Used for Workflow Commissioning PT, Commissioning and Annual QA RT, Annual QA PT, Monthly QA RT, Monthly QA PT