Skip to content

Quality Assurance & Dosimetry

FC23-C Ionization Chamber

FC23-C is a shortened farmer-type ionization chamber for absolute dosimetry of photon, electron, and proton beams at therapy-level dose rates. FC23-C measures absorbed dose to water, air kerma, and exposure in air, depending on radiation type and calibration.

FC23-C Ionization Chamber

FC23-C is a shortened farmer-type ionization chamber for absolute dosimetry of photon, electron, and proton beams at therapy-level dose rates. FC23-C measures absorbed dose to water, air kerma, and exposure in air, depending on radiation type and calibration. Designed for medical accelerators, it also works with X-ray beams (70–280 kV), gamma radiation (¹³⁷Cs, ⁶⁰Co), and proton beams (50–250 MeV).

Used for: Detectors RT, Monthly QA RT, Commissioning and Annual QA RT

Features

Inner dimensions

  • Sensitive volume (nominal) 0.23 cm³
  • Total sensitive length: 9.0 mm
  • Total cylinder length: 6.0 mm
  • Inner diameter of outer electrode: 6.2 mm
  • Wall thickness of outer electrode: 0.4 mm (0.056 g/cm²)
  • Diameter of inner electrode: 1.0 mm
  • Length of inner electrode 7.5 mm

Operational characteristics

  • Polarizing voltage: ±300 V (max. ±500 V)
  • Typical leakage current: 3 fA
  • Recommended pre-irradiation: 5 Gy
  • Typical sensitivity: 7.2 nC/Gy
  • Guard potential: ±300 V (max. ±500 V)
  • Temperature range: 15 °C – 35 °C
  • Relative humidity range: 20 % – 80 %
  • FC23 Ion Chamber with Chamber Cap
  • FC Ion Chamber with Chamber Cap in Large Universal Chamber Holder
  • FC Ion Chamber in Large Universal Chamber Holder

Technical specifications

  • FC65-G — Cavity volume (cm³): 0.65; Cavity length (mm): 23.1; Cavity radius (mm): 3.1; Wall material: Graphite; Wall thickness (g/cm²): 0.073; Central electrode material: Aluminium; Waterproof: Y
  • FC65-P — Cavity volume (cm³): 0.65; Cavity length (mm): 23.1; Cavity radius (mm): 3.1; Wall material: POM; Wall thickness (g/cm²): 0.057; Central electrode material: Aluminium; Waterproof: Y
  • FC23-C — Cavity volume (cm³): 0.23; Cavity length (mm): 8.8; Cavity radius (mm): 3.1; Wall material: C552; Wall thickness (g/cm²): 0.07; Central electrode material: C552; Waterproof: Y

Technical data is subject to change without prior notice.

In this division

DOSE-X

Next-Generation Reference Class Electrometer DOSE-X is a portable reference class electrometer exceeding IEC 60731 standards. Discover your perfect solution for conducting reference measurements according to DIN 6800, AAPM TG51, and IAEA TRS398. Compatible with Machine/Modality/Application: Carbon Compatible , C-arm linacs , Robotic linac , Cyclotrons , Bore-type linacs , IMRT / VMAT , Proton Therapy , Helical linac , Small Field Dosimetry , Synchro-Cyclotron , Ultra-high Dose Rate (UHDR) , TG106, TG142, TG148, TG224, TG51 Used for Workflow Commissioning PT, Commissioning and Annual QA RT, Annual QA PT, Monthly QA RT, Monthly QA PT

Talk to our team

Questions about a system, a quotation or after-sales support? Call or email us and the right person will get back to you.