Skip to content

Quality Assurance & Dosimetry

CC25 Ionization Chamber

CC25 is intended for absolute and relative dosimetry of photon, electron, and proton beams in radiotherapy. Measurements can be performed in air, water phantoms, or solid phantoms. CC25 is a high spatial resolution, small volume chamber.

CC08 Ionization Chamber

CC25 is intended for absolute and relative dosimetry of photon, electron, and proton beams in radiotherapy. Measurements can be performed in air, water phantoms, or solid phantoms. CC25 is a high spatial resolution, small volume chamber. It is waterproof, vented through a waterproof sleeve, and fully guarded.

Used for: Detectors RT

Features

Inner dimensions

  • Active volume: 0.25 cm³
  • Total active length: 10 mm
  • Cylinder length: 7 mm
  • Inner diameter of outer electrode: 6.0 mm
  • Wall thickness: 0.4 mm (70 mg/cm²)
  • Diameter of inner electrode: 1.0 mm
  • Length of inner electrode: 7.5 mm

Operational characteristics

  • Polarizing voltage: ±300 V (max. ±500 V)
  • Typical leakage current: 5 fA
  • Recommended pre-irradiation: 5 Gy
  • Typical sensitivity: 7.5 nC/Gy
  • Guard potential: ±300 V (max. ±500 V)
  • Temperature range: 15 °C – 35 °C
  • Relative humidity range: 20 % – 80 %

Technical specifications

  • CC04 — Cavity Volume (cm³): 0.04; Cavity Length (mm): 3.6; Cavity Radius (mm): 2; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
  • CC08 — Cavity Volume (cm³): 0.08; Cavity Length (mm): 4; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
  • CC13 — Cavity Volume (cm³): 0.13; Cavity Length (mm): 5.8; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
  • CC25 — Cavity Volume (cm³): 0.25; Cavity Length (mm): 10; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y

Technical data is subject to change without prior notice.

In this division

DOSE-X

Next-Generation Reference Class Electrometer DOSE-X is a portable reference class electrometer exceeding IEC 60731 standards. Discover your perfect solution for conducting reference measurements according to DIN 6800, AAPM TG51, and IAEA TRS398. Compatible with Machine/Modality/Application: Carbon Compatible , C-arm linacs , Robotic linac , Cyclotrons , Bore-type linacs , IMRT / VMAT , Proton Therapy , Helical linac , Small Field Dosimetry , Synchro-Cyclotron , Ultra-high Dose Rate (UHDR) , TG106, TG142, TG148, TG224, TG51 Used for Workflow Commissioning PT, Commissioning and Annual QA RT, Annual QA PT, Monthly QA RT, Monthly QA PT

Talk to our team

Questions about a system, a quotation or after-sales support? Call or email us and the right person will get back to you.