Skip to content

Quality Assurance & Dosimetry

CC13 Ionization Chamber

CC13 is the standard chamber for clinical use in water phantoms and for measurements of output factors. CC13 is intended for absolute and relative dosimetry of photon, electron, and proton beams in radiotherapy. Measurements can be performed in air, water phantoms, or solid phantoms.

CC08 Ionization Chamber

CC13 is the standard chamber for clinical use in water phantoms and for measurements of output factors. CC13 is intended for absolute and relative dosimetry of photon, electron, and proton beams in radiotherapy. Measurements can be performed in air, water phantoms, or solid phantoms.

Compatible with: Small Field Dosimetry

Used for: Commissioning and Annual QA RT, Detectors RT

Features

Inner dimensions

  • Active volume: 0.13 cm³
  • Total active length: 5.8 mm
  • Cylinder length: 2.8 mm
  • Inner diameter of outer electrode: 6.0 mm
  • Wall thickness: 0.4 mm (70 mg/cm²)
  • Diameter of inner electrode: 1.0 mm
  • Length of inner electrode: 3.3 mm

Operational characteristics

  • Polarizing voltage: ±300 V (max. ±500 V)
  • Typical leakage current: 3 fA
  • Recommended pre-irradiation: 5 Gy
  • Typical sensitivity: 3.6 nC/Gy
  • Guard potential: ±300 V (max. ±500 V)
  • Temperature range: 15 °C – 35 °C
  • Relative humidity range: 20 % – 80 %
  • Compact Ion Chamber in Small Universal Chamber Holder
  • Compact Ion Chamber in Large Universal Chamber Holder

Technical specifications

  • CC04 — Cavity Volume (cm³): 0.04; Cavity Length (mm): 3.6; Cavity Radius (mm): 2; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
  • CC08 — Cavity Volume (cm³): 0.08; Cavity Length (mm): 4; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
  • CC13 — Cavity Volume (cm³): 0.13; Cavity Length (mm): 5.8; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
  • CC25 — Cavity Volume (cm³): 0.25; Cavity Length (mm): 10; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y

Technical data is subject to change without prior notice.

In this division

DOSE-X

Next-Generation Reference Class Electrometer DOSE-X is a portable reference class electrometer exceeding IEC 60731 standards. Discover your perfect solution for conducting reference measurements according to DIN 6800, AAPM TG51, and IAEA TRS398. Compatible with Machine/Modality/Application: Carbon Compatible , C-arm linacs , Robotic linac , Cyclotrons , Bore-type linacs , IMRT / VMAT , Proton Therapy , Helical linac , Small Field Dosimetry , Synchro-Cyclotron , Ultra-high Dose Rate (UHDR) , TG106, TG142, TG148, TG224, TG51 Used for Workflow Commissioning PT, Commissioning and Annual QA RT, Annual QA PT, Monthly QA RT, Monthly QA PT