CC08 is used for customized applications during manufacturing and installation of linear accelerators (e.g., “Buddelship”). CC08 is a high spatial resolution, small volume chamber. It is waterproof, vented through a waterproof sleeve, and fully guarded.
Compatible with: TG106
Used for: Commissioning and Annual QA RT, Detectors RT
Features
Inner dimensions
- Active volume: 0.08 cm³
- Total active length: 4.0 mm
- Cylinder length: 1.0 mm
- Inner diameter of outer electrode: 6.0 mm
- Wall thickness 0.4 mm (70 mg/cm²)
- Diameter of inner electrode: 1.0 mm
- Length of inner electrode: 1.5 mm
Operational characteristics
- Polarizing voltage: ±300 V (max. ±500 V)
- Typical leakage current: 3 fA
- Recommended pre-irradiation: 5 Gy
- Typical sensitivity: 1.9 nC/Gy
- Guard potential: ±300 V (max. ±500 V)
- Temperature range: 15 °C – 35 °C
- Relative humidity range: 20 % – 80 %
Technical specifications
- CC04 — Cavity Volume (cm³): 0.04; Cavity Length (mm): 3.6; Cavity Radius (mm): 2; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
- CC08 — Cavity Volume (cm³): 0.08; Cavity Length (mm): 4; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
- CC13 — Cavity Volume (cm³): 0.13; Cavity Length (mm): 5.8; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
- CC25 — Cavity Volume (cm³): 0.25; Cavity Length (mm): 10; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
Technical data is subject to change without prior notice.