CC04 is the conventional ionization chamber for measurements of small fields and ranges with high dose gradients, e.g., stereotactic fields. CC04 is intended for absolute and relative dosimetry of photon, electron, and proton beams in radiotherapy. Measurements can be performed in air, water phantoms, or solid phantoms. CC04 is a high spatial resolution, small volume chamber. It is waterproof, vented through a waterproof sleeve, and fully guarded.
Compatible with: Small Field Dosimetry, TG106
Used for: Commissioning and Annual QA RT, Detectors RT
- CC04 in Small Universal Holder
- CC04 in Large Universal Holder
Features
Inner dimensions
- Active volume: 0.04 cm³
- Total active length: 3.6 mm
- Cylinder length: 1.6 mm
- Inner diameter of outer electrode: 4.0 mm
- Wall thickness: 0.4 mm (70 mg/cm²)
- Diameter of inner electrode: 1.0 mm
- Length of inner electrode: 2.1 mm
Operational characteristics
- Polarizing voltage: ±300 V (max. ±500 V)
- Typical leakage current: 3 fA
- Recommended pre-irradiation: 5 Gy
- Typical sensitivity: 1 nC/Gy
- Guard potential: ±300 V (max. ±500 V)
- Temperature range: 15 °C – 35 °C
- Relative humidity range: 20 % – 80 %
Technical specifications
- CC04 — Cavity Volume (cm³): 0.04; Cavity Length (mm): 3.6; Cavity Radius (mm): 2; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
- CC08 — Cavity Volume (cm³): 0.08; Cavity Length (mm): 4; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
- CC13 — Cavity Volume (cm³): 0.13; Cavity Length (mm): 5.8; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
- CC25 — Cavity Volume (cm³): 0.25; Cavity Length (mm): 10; Cavity Radius (mm): 3; Wall Material: C552; Wall Thickness (g/cm²): 0.07; Central Electrode Material: C552; Water-proof: Y
Technical data is subject to change without prior notice.