The 10XF-3CT ionization chamber is the chamber of choice for measurements on CT scanners with your MagicMax .
Key Benefits
- Flat energy response covering the total active chamber length
- High confidence with individual factory calibration sheet
- Air ionization chamber
- Homogeneous sensitivity over the entire active length of the chamber
- Provided with protective shielding (guarded)
- Delivered with an individual calibration document
Technical Specifications
Material
- External electrode: C552
- Inner electrode: C552
- Connector: 7-pin multi plug connector
- Cable: 2 m flexible, low noise
Size
- Active volume: 3 cm³
- Total active length: 100 mm
- Outer diameter of the external diode: 9 mm
Operating Data
- Leakage current: < ± 4 x 10 -15 A
- Measuring range for CT beam quality: RQR 8 - RQR 10; 100 kV - 150 kV
- RQR 8 - RQR 10: 100 kV - 150 kV
- RQR 8 - RQR 9: 100 kV - 120 kV
- Calibration factor (typical): N D,K = 72 mGy cm/nC (120 kV/4.5 mm Al HVL)
- Sensitivity: 1.11 nC/ Gy
- Dose measuring range: 0.032 mGy – 15 Gy
- Dose rate measuring range: 0.32 mGy/s – 0.05 Gy/s
- Dose length measuring range: 0.32 mGycm – 132 Gycm
- Dose rate length range: 3.2 mGycm/s – 0.5 Gycm/s
- Uncertainty: < 5 %
- Calibration reference: RQT 9: in the chamber axis at the center of the cavity volume
- Beam incidence direction: the central beam axis is perpendicular to the chamber axis